  • 新款粘度计.png

SSR Series Digital Rotary Viscometer for Laboratory Viscosity Measurement

Number:

DH-RV-SSR

Activity price:

¥

Market price:

¥


Details


Related Products

SSR Series Digital Rotary Viscometer for Laboratory Viscosity Measurement

 

Feature:

1. Accurate viscosity measurement value: each range is automatically calibrated by computer, with high accuracy and small error;

2. Front-level instrument: the level adjustment is intuitive and convenient;

3. Optional Pt100 temperature probe: wide temperature measurement range, from-20 to 300℃, temperature measurement accuracy of 0.1℃;

4. Optional enhanced ultra-low viscosity adapter ULR / URL PLUS;

5. The sample volume of each measurement is only 7-11ml;

6. Rich in optional accessories: constant temperature bath, printer, standard viscosity sample (standard silicone oil), etc.;

7. Automatic scanning, timing measurement and other functions;

8. Automatic prompt function of viscosity measurement stability;

9. Display the shear rate and shear stress;

10. Viscosity unit switching (1Pa.s=1000mPa.s; 1P=100mPa.s; 1cP=1mPa.s);

11. Temperature unit switching: Celsius, Fahrenheit;

12. Connect to the printer and the computer

13. Switch between Chinese and English operating systems

 

The torque of the gossamer, the shape of the rotor, the size of the rotation speed, the range of viscosity measurement are the same as the same type of viscometer, and the measured viscosity is very comparable with the imported similar instruments.

LV series viscometer measures low viscosity samples, the main measured samples are: solvent type adhesive, latex preparation, oil, oil, chemical reagents, paint, paint, paint, cosmetics, drugs, daily chemicals, photosensitive resin, hot paraffin, polymer solution, ink, rubber solution, fruit juice, solvent, etc.

 

The RV series viscometer is suitable for medium viscosity fluid, and the main measured samples are:

Paint, asphalt (SHRP), paper coating, ceramics, pulp, frost, like objects, plastic sols, daily chemicals, starch, food products, surface coatings, resin, gum, toothpaste, silk mark ink, bright agent, organic solvents

 

Main technical parameters:

Model

DH-LV-SSR

Display mode

Digital display

Speed(r/min)

0.3 – 100  Total 37

0.3、0.5、0.6、1、1.5、2、2.5、3、4、5、6、7、8、9、10、12、14、15、16、18、20、25、30、35、40、45、50、55、60、65、70、75、80、85、90、95、100RPM

Measuring range

mPa.s

1 - 1.6M

ULR: 1 - 1K

18 rotor:3 - 10K

25 rotor:480 - 1.6M

31 rotor:30 - 100K

34 rotor:60 - 200K

Remarks: K=1000,M=1000000

Sample size

ULR: 21ml

18,25,31,34 rotor:7-11ml

Measurement error

±1%  (Newtonian liquid)

Repetitive error

±0.3%  (Newtonian liquid)

The shear response

/ shear rate is displayed

Standard configuration

Timed function

Standard configuration

Temperature measurement function

Standard temperature probe interface

(optional temperature probe PT 100)

Automatic scanning function

Automatically scan and recommend a preferential combination of the rotor and the rotation speed

Maximum measurement range

Automatically displays the measurable viscosity range of the selected rotor combination and rotational speed

Put a seal on

Data and curve can be printed

(standard with printing interface, buy printer)

Data output interface

RS232: Connect printer (dedicated);

RS232: connect computer (dedicated)

Thermostatic parts

Option parts (constant temperature circulating water jacket, special constant temperature tank for viscosity meter)

Working power supply

Wide voltage operation (110V / 60Hz or 220V / 50Hz)

Outline dimension

300 × 300 × 450 (mm)

 

Model

DH-RV-SSR

DH-HA-SSR

DH-HB-SSR

Display mode

Digital display

 

Speed(r/min)

0.3 – 100  Total 37

0.3、0.5、0.6、1、1.5、2、2.5、3、4、5、6、7、8、9、10、12、14、15、16、18、20、25、30、35、40、45、50、55、60、65、70、75、80、85、90、95、100RPM

Measuring range

mPa.s

6.4 - 3.3M

ULR: 6.4 - 1K

21 rotor:50-167K

27 rotor:250-834K

28 rotor:500-1.7M

29 rotor:1K-3.3M

12.8 - 6.6M

ULR: 12.8 - 1K

21 rotor:100-333K

27 rotor:500-1.7M

28 rotor:1K-3.3M

29 rotor:2K-6.6M

51.2 - 26.6M

ULR: 51.2-2k

21 rotor:400-1.3M

27 rotor:2K-6.7M

28 rotor:4K-13.3M

29 rotor:8K-26.6M

Remarks:K=1000,M=1000000(million)

Sample size

ULR: 21ml

21,27,28,29 rotor: 8 - 13ml

Measurement error

±1%  (Newtonian liquid)

Repetitive error

±0.3%  (Newtonian liquid)

The shear response / shear rate is displayed

standard configuration

Timed function

standard configuration

Temperature measurement function

Standard temperature probe interface

(optional temperature probe PT 100)

Automatic scanning function

Automatically scan and recommend a preferential combination of the rotor and the rotation speed

Maximum measurement range

Automatically displays the measurable viscosity range of the selected rotor combination and rotational speed

Put a seal on

Data and curve can be printed

 (standard with printing interface, buy printer)

Data output interface

RS232: Connect printer (dedicated);

RS232: connect computer (dedicated)

Thermostatic parts

Option parts (constant temperature circulating water jacket, special constant temperature tank for viscosity meter)

Working power supply

Wide voltage operation (110V / 60Hz or 220V / 50Hz)

Outline dimension

300 × 300 × 450 (mm)

 

 

Select accessories:

 

1. Constant temperature circulating water jacket

If you want to control the temperature of the tested samples, it is necessary to choose a constant temperature circulating water jacket and connect the constant temperature tank with a hose. The temperature controlled water flow in the temperature tank passes through the water jacket to transfer the temperature to the sample.

 

 

2. DC0506W viscosity meter

 

Using high performance monolithic microcomputer control, self-setting PID adjustment, imported from Japan platinum resistance, imported nbraco (embrsco) refrigeration compressor, fast cooling, accurate temperature control. Using large screen backlight LCD screen, the parameters are clear, with broken protection constant temperature level protection device, can effectively prevent the instrument out of control and heating pipe empty burning and other phenomena.

 

The instrument works stable and reliable and easy to operate.

In the full range of temperature control, the linearity is very

good,0.1 DCW (N) series with temperature control

The degree error is less than 0.2℃, and the temperature

fluctuation is less than ± 0.05;

Temperature control range: -5 to 100℃

The above types of viscometer can also be selected with temperature control device and heating furnace combination Chen high temperature Brell viscometer, used to measure the high temperature melting samples, such as paraffin, polyethylene wax, asphalt, hot melt glue, plastic and other samples.